Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8000A: CAL ADJ /5700 DATE: 18-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 93 CONFIGURATION: Philips PM 6666 CONFIGURATION: Fluke 5700A ============================================================================= # # Source: # Fluke 8000A instruction manual (PN 347906). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- N J W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP Remove the screw located just below the AC line 1.003 DISP jack. Slide the instrument out of the case. 1.004 DISP Connect the UUT to an ac power source. 1.004 DISP Turn the UUT on. 1.004 DISP Warm-up time: 1 hour. 1.004 DISP Ambient temperature: 22C - 25C. 1.004 DISP Relative humidity: less than 70%. 1.005 HEAD Period Adjustment 1.006 DISP Connect the PM 6666 and UUT as follows: 1.006 DISP [32] PM 6666 INPUT A HI to UUT TP5 1.006 DISP [32] PM 6666 INPUA A LO to UUT INPUT LO 1.007 MESS Adjust R20 for a time period of 100ms (+/-5us). 1.007 MESS Variations of the time period must be <= 15us. 1.008 6666 100mT 2V +D PA S 2W 1.009 MESS 1.010 DISP Disconnect the PM 6666 from the UUT. 1.011 HEAD ZERO OFFSET ADJUSTMENT 1.012 DISP Connect the 5700A and the UUT as follows. 1.012 DISP [32] 5700A OUTPUT HI to UUT V-Ohm 1.012 DISP [32] 5700A OUTPUT LO to UUT COMMON 1.012 DISP [32] 5700A SENSE HI to UUT V-Ohm 1.012 DISP [32] 5700A SENSE LO to UUT COMMON 1.013 RNG 200mV 1.014 5700 0mV S 2W 1.015 DISP Adjust R15 for a UUT display of 00.0+/-00.1. 1.016 5700 * S 1.017 OPBR Does the UUT read less than or equal to 01.0? 1.018 JMPF 1.015 1.019 HEAD TURN-OVER ERROR ADJUSTMENT 1.020 5700 190mV S 2W 1.021 DISP Adjust R23 for a UUT reading of 190.0mV. 1.022 5700 -190mV S 2W 1.023 DISP Adjust R15 for a UUT reading of 190.0+/-0.1mV. 1.024 HEAD RANGE ADJUSTMENTS (DC Volts) 1.025 5700 190mV S 2W 1.026 DISP Adjust R23 for a UUT reading of 190.0+/-0.3mV. 1.027 5700 * S 1.028 RNG 2V 1.029 5700 1.9V S 2W 1.030 DISP Adjust R25 for a UUT reading of 1.900+/-0.003V. 1.031 5700 * S 1.032 HEAD RANGE ADJUSTMENTS (Ohms) 1.033 RNG 20MZ 1.034 5700 19MZ S 2W 1.035 DISP Adjust R55 for a UUT reading of 19.00+/-0.11 MOhm. 1.036 5700 * S 1.037 RNG 20kZ 1.038 5700 19kZ S CW 1.039 DISP Adjust R12 for a UUT reading of 19.00+/-0.05 kOhm. 1.040 5700 * S 1.041 HEAD RANGE ADJUSTMENTS (High Freq. AC) 1.042 RNG 20V AC 1.043 5700 19V 20kH S 2W 1.044 DISP Adjust C3 for a UUT reading of 19.00+/-0.21V. 1.045 END